Grab Sample Conditioning Panel SL200H Special

https://industrial.intrial.com.pe/web/image/product.template/51407/image_1920?unique=042564e

SL200H-SPECIAL<br />
Sample: Boiler Water 120C° at 60 BARG<br />
Conditioning Panel for On-Line Analysis<br />
Consisting of the following components:<br />
* 1/4 in. Sample Inlet Isolation Valve<br />
* 1/4 in. High Pressure (Sample Inlet) Blowdown Valve<br />
* Primary Sample Cooler, Sentry Model TLR4225<br />
* 3-Way Cooling Water Inlet Valve, with drain<br />
* Globe Valve on the Cooling Water Outlet<br />
* Cooling Water Sight Gauge<br />
* Cooling Water Valve Materials are Stainless Steel<br />
* Sentry VREL High Pressure Reducing Valve<br />
* Sentry TSV with a Setpoint of 120°F(49°C)<br />
* Total Flow Indicator, 0-1600 cc/min<br />
* Temperature Gauge, 0-250°F/-20-120°C<br />
* Sentry BPRV with a Setpoint of 20 psig/1.4 barg<br />
* 304 SS Sample Sink<br />
All components are factory mounted on a 12 in. W x 36 in. H, T304SS backplate.<br />
All sample tubing throughout panel is SA249, T316 SS.<br /><br />
QA Documents Provided:<br />
* QAF 1050 Manufacture Certificate for Testing and Material – Single Line<br />
* QAF 1030 Product Inspection Checklist - Includes Hydrotest Results